Verifying Thermocouple Simulation (Sourcing); Verifying Thermocouple Measurement - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual

3-30. Verifying Thermocouple Simulation (Sourcing)

Verify that the 5520A performance is within the limits in Table 3-30. Use the HP3458A
DMM as the measurement device. Use copper connectors and copper wires.
TC Type
10 V/ C

3-31. Verifying Thermocouple Measurement

Verify that the 5520A performance is within the limits in Table 3-31. Use a Fluke 5500A
Calibrator or similar instrument as the millivolt source, connected in parallel with an
HP3458A DMM. At each verification point, use the 5500A error mode controls to adjust
the calibrator output for a nominal reading on the DMM. Use copper connectors and
copper wires.
TC Type
10 V/ C
3-52
Table 3-30. Verification Tests for Thermocouple Simulation
Output, C
0.00 C (0.0000 mV)
100.00 C (1.0000 mV)
-100.00 C (-1.0000 mV)
1000.00 C (10.0000 mV)
-1000.00 C (-10.0000 mV)
10000.00 C (100.0000 mV)
-10000.00 C (-100.0000 mV)
Table 3-31. Verification Tests for Thermocouple Measurement
Input, mV
0.00 C (0.0000 mV)
10000.00 C (100.0000 mV)
-10000.00 C (-100.0000 mV)
30000.00 C (300.0000 mV)
-30000.00 C (-300.0000 mV)
Lower Limit, mV
Upper Limit, mV
-0.0030
0.99696
-1.00304
9.99660
-10.0034
99.9930
-100.0070
Lower Limit, C
-0.30
9999.30
-10000.70
29998.50
-30001.50
0.0030
1.00304
-.99696
10.00340
-9.9966
100.0070
-99.9930
Upper Limit, C
0.30
10000.70
-9999.30
30001.50
-29998.50