Leveled Sine Wave Harmonics Verification - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Table 6-53. Leveled Sine Wave Frequency Verification
Calibrator Mainframe
Frequency
(Output @ 5.5 V p-p)
50 kHz
500 kHz
5 MHz
50 MHz
500 MHz

6-126. Leveled Sine Wave Harmonics Verification

This procedure uses the following equipment:
Hewlett-Packard 8590A Spectrum Analyzer
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC300
Refer to Figure 6-24 for proper setup connections.
HP 8590A
BNC(F)
to Type N (M)
Adapter
Figure 6-24. Leveled Sine Wave Harmonics Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Levsine menu on the display.
Then follow these steps to verify the leveled sine wave harmonics.
1. Using the BNC cable and BNC(f) to Type N(m) adapter, connect the SCOPE
connector on the Calibrator Mainframe to the HP 8590A.
2. Program the Calibrator Mainframe to 5.5 V p-p at each frequency listed in Table 6-
54. Press
PM 6680 Settings
Channel
Filter
A
On
A
Off
A
Off
A
Off
C
Off
SC300
on the Calibrator Mainframe to activate the output.
PM 6680 Reading
(Frequency)
Cable
SC300 Option
Verification
Tolerance
1.25 Hz
12.5 Hz
125.0 Hz
1250 Hz
12500 Hz
5520A-SC300
5520A
CALIBRATOR
NORMAL
AUX
SCOPE
V, , ,RTD
A, -SENSE, AUX V
OUT
HI
1000V
20V
RMS
1V PK
RMS
MAX
MAX
MAX
150V
LO
PK
MAX
TRIG
20V
RMS
MAX
20V
PK
20A
SHELLS
GUARD
NOT
MAX
GROUNDED
TC
20V PK MAX
20V PK MAX
yg125f.eps
6
6-97