Fluke 5520A Service Manual page 215

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Table 6-41. SC300 Calibration and Verification Equipment (cont.)
Instrument
Frequency
PM 6680 with option (PM 9621, PM 9624, or
Counter
PM 9625) and (PM 9678)
Adapter
Pomona #3288
BNC Cable
(supplied with SC300)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
AC Measurement
Fluke 5790A
Standard
with -03 option
Adapter
Pomona #3288
BNC Cable
(supplied with SC300)
Spectrum Analyzer
Adapter
BNC Cable
Edge Frequency, AC Voltage Frequency Verification
Frequency Counter
BNC Cable
Frequency Counter
BNC Cable
Leveled Sine Wave Flatness (High Frequency) Calibration and Verification
Power Meter
Hewlett-Packard 437B
Power Sensor
Hewlett-Packard 8482A
Power Sensor
Hewlett-Packard 8481D
30 dB
Hewlett-Packard
Reference
11708A
Attenuator
(supplied with HP
8481D)
Adapter
Hewlett-Packard
PN 1250-1474
BNC Cable
(supplied with SC300)
Model
Leveled Sine Wave Frequency Verification
Range
Frequency
BNC(f) to Type N(m)
Leveled Sine Wave Harmonics Verification
HP 8590A
Pomona #3288
(supplied with SC300)
PM 6680 with option (PM
9678)
(supplied with SC300)
Edge Duty Cycle
PM 6680
(supplied with SC300)
Range
Frequency
Range
Frequency
Range
Frequency
Range
Frequency
BNC(f) to Type N(f)
Equipment Required for Calibration and Verification
Minimum Use Specifications
50 kHz to 350 MHz, < 1.6 ppm
uncertainty
BNC(f) to Type N(m)
5 mV p-p to 5.5 V p-p
50 kHz to 10 MHz
BNC(f) to Type N(m)
20 ms to 150 ns, 10 Hz to 10 MHz: < 1.6 ppm
uncertainty
-42 to +5.6 dBm
10 - 300 MHz
-20 to +19 dBm
10 - 300 MHz
-42 to -20 dBm
10 - 300 MHz
30 dB
50 MHz
SC300 Option
6
6-75