Fluke 5520A Service Manual page 191

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Calibrator
Calibrator
Mainframe
Mainframe
output
Wave Type
(@ 10 kHz)
square
1.8 mV
square
11.9 mV
square
21.9 mV
square
22.0 mV
square
56.0 mV
square
89.9 mV
square
90 mV
square
155 mV
square
219 mV
square
220 mV
square
560 mV
square
899 mV
square
0.90 V
square
3.75 V
square
6.59 V
square
6.6 V
square
30.8 V
square
55.0 V
sine
1.8 mV
sine
21.9 mV
sine
89.9 mV
sine
219 mV
sine
899 mV
sine
6.59 V
sine
55 V
triangle
1.8 mV
triangle
21.9 mV
triangle
89.9 mV
triangle
219 mV
triangle
899 mV
triangle
6.59 V
triangle
55 V
Table 6-35. Wave Generator Verification at 1 M
5790A
Conversion
Reading
(V rms)
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
5790A Reading x
Conversion Factor
Factor
(V p-p)
SC600 Option
6
Verification
Tolerance
(V p-p)
0.000154 V
0.000457 V
0.00075 V
0.00076 V
0.00178 V
0.002797 V
0.0028 V
0.00475 V
0.00667 V
0.0067 V
0.0169 V
0.02707 V
0.0271 V
0.1126 V
0.1978 V
0.1981 V
0.9241 V
1.6501 V
0.000154 V
0.000757 V
0.002797 V
0.00667 V
0.02707 V
0.1978 V
1.6501 V
0.000154 V
0.000757 V
0.002797 V
0.00667 V
0.02707 V
0.1978 V
1.6501 V
6-51