Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual

1-8. Specifications

The following tables list the 5520A specifications. All specifications are valid after
allowing a warm-up period of 30 minutes, or twice the time the 5520A has been turned
off. (For example, if the 5520A has been turned off for 5 minutes, the warm-up period is
10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures
outside of t
temperature coefficient as stated in the General Specifications must be applied.
The specifications also assume the Calibrator is zeroed every seven days or whenever the
ambient temperature changes more than 5 C. The tightest ohms specifications are
maintained with a zero cal every 12 hours within 1 C of use.
Also see additional specifications later in this chapter for information on extended
specifications for ac voltage and current. The dimensional outline for the 5520A
Calibrator is shown in Figure 1-3.
5520A CALIBRATOR
NORMAL
V, , ,RTD
A, -SENSE, AUX V
HI
LO
GUARD
20V PK MAX
1-6
5 C (t
is the ambient temperature when the 5520A was calibrated), the
cal
cal
43.2 cm (17 in)
AUX
SCOPE
OUT
STBY
OPR
EARTH
EXGRD
SCOPE
7
8
9
m
TRIG
n
4
5
6
k
p
20A
1
2
3
M
/ +
0
SHIFT
TC
20V PK MAX
47.0 cm (18.5 in)
Figure 1-3. 5520A Calibrator Dimensional Outline
PREV
MENU
dBm
sec
EDIT
SETUP
RESET
V
Hz
FIELD
W
¡F
NEW
CE
A
¡C
REF
MEAS
MORE
F
TC
MODES
MULT
DIV
ENTER
x
17.8 cm
(7 in)
POWER
I
O
6.4 cm
(2.5 in)
For Cable
Access
nn032f.eps

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents