AC Voltage (Sine Wave) Specifications (cont)
Range
Frequency
10 mV to
10 Hz to 20 Hz
329.999 mV
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 Hz to 30 kHz
0.33 V to
10 Hz to 20 Hz
3.29999 V
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30 kHz
3.3 V to 5 V
10 Hz to 20 Hz
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
[1] There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is < 5 mΩ for outputs ≥0.33 V. The AUX output
resistance is < 1Ω. The maximum load capacitance is 500 pF, subject to the maximum burden
current
limits.
AUX (Auxiliary Output) [dual output mode only] [1]
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µV)
90 days
0.15 + 370
0.2 + 370
0.08 + 370
0.1 + 370
0.08 + 370
0.1 + 370
0.15 + 450
0.2 + 450
0.3 + 450
0.4 + 450
5.0 + 900
4.0 + 900
0.15 + 450
0.2 + 450
0.08 + 450
0.1 + 450
0.07 + 450
0.09 +
450
0.15 + 1400
0.2 +
1400
0.3 + 1400
0.4 +
1400
4.0 + 2800
5.0 +
2800
0.15 + 450
0.2 + 450
0.08 + 450
0.1 + 450
0.07 + 450
0.09 +
450
0.15 + 1400
0.2 +
1400
0.3 +1400
0.4 +
1400
Introduction and Specifications
Resolution
Burden
1 year
1 µV
5 mA
10 µV
5 mA
100 µV
5 mA
Specifications
Max Distortion
and Noise
Max
10 Hz to
100 kHz
Bandwidth
± (% output +
floor)
0.2 + 200 µV
0.06 + 200 µV
0.08 + 200 µV
0.3 + 200 µV
0.6 + 200 µV
1 + 200 µV
0.2 + 200 µV
0.06 + 200 µV
0.08 + 200 µV
0.3 + 200 µV
0.6 + 200 µV
1 + 200 µV
0.2 + 200 µV
0.06 + 200 µV
0.08 + 200 µV
0.3 + 200 µV
0.6 + 200 µV
1