Leveled Sine Wave Specifications; Time Marker Specifications; Wave Generator Specifications - Fluke 5502A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

Leveled Sine Wave Specifications

Leveled Sine Wave
Qualities into 50 Ω
Amplitude Qualities (for measuring oscilloscope bandwidth)
Range (p-p)
Resolution
Adjustment Range
1-Year Absolute
Uncertainty, tcal ±5 °C
Flatness (relative to
50 kHz)
Short-Term Amplitude
Stability
Frequency Qualities
Resolution
1-Year Absolute
Uncertainty, tcal ±5 °C
Distortion Qualities
2nd Harmonic
3rd and Higher Harmonics
Within 1 hour after reference amplitude setting, provided temperature varies no more than ±5 °C.
[1]
[2]
With REF CLK set to ext, the frequency uncertainty of the Leveled Sine Wave is the uncertainty of the external 10 MHz clock
±0.3 Hz/gate time.

Time Marker Specifications

Time Maker into 50 Ω
1-Year Absolute
±(25 + t *1000)
Uncertainty at Cardinal
ppm
Points, tcal ±5 °C
Wave Shape
spike or square
Typical Output Level
>1 V p-p
Typical Jitter (rms)
<10 ppm
Sequence
5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
[3]
At least ±10 % around each sequence value indicated above.
Adjustment Range
Amplitude Resolution
4 digits
[1]
t is the time in seconds.
[2]
Typical rise time of square wave and 20 %-pulse (20 % duty cycle pulse) is < 1.5 ns.
Time marker uncertainty is ±50 ppm away from the cardinal points.
[3]

Wave Generator Specifications

Wave Generator Qualities
Amplitude
Range
1-Year Absolute Uncertainty, tcal ±5 °C,
10 Hz to 10 kHz
Sequence
Typical DC Offset Range
Frequency
Range
Resolution
1-Year Absolute Uncertainty, tcal ±5 °C
[1]
The DC offset plus the wave signal must not exceed 30 V rms.
50 kHz (reference)
50 kHz to 100 MHz
±(2 % of output
±(3.5 % of output
+ 300 μV)
+ 300 μV)
not applicable
±(1.5 % of output
+ 100 μV)
1 kHz
20 ms to
5 s to 50 ms
100 ns
±2.5 ppm
[1]
spike, square,
or 20 %-pulse
[2]
[2]
>1 V p-p
<1 ppm
into 1 MΩ:
into 50 Ω:
±(3 % of p-p output + 100 µV)
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
0 to ± (≥40 % of p-p amplitude)
10 Hz to 100 kHz
4 or 5 digits depending upon frequency
±(25 ppm + 15 mHz)
SC600 Oscilloscope Calibration Option
SC600 Oscilloscope Calibration Option Specifications
Frequency Range
100 to 300 MHz
5 mV to 5.5 V
<100 mV: 3 digits
≥100 mV: 4 digits
continuously adjustable
±(4 % of output
+ 300 μV)
±(2 % of output
+ 100 μV)
[1]
≤ 1 %
[2]
±2.5 ppm
≤ -33 dBc
≤ -38 dBc
50 to 20 ns
10 ns
±2.5 ppm
±2.5 ppm
spike or square
square or sine
[2]
>1 V p-p
>1 V p-p
<1 ppm
<1 ppm
Square Wave, Sine Wave, and Triangle Wave
into 50 Ω or 1 MΩ
1.8 mV to 55 V p-p
1.8 mV to 2.5 V p-p
[1]
300 to 600 MHz
±(6 % of output
+ 300 μV)
±(4 % of output
+ 100 μV)
10 kHz
5 to 2 ns
±2.5 ppm
sine
[2]
>1 V p-p
<1 ppm
9
9-5