Fluke 5520A Service Manual page 167

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

BNC(F)
Figure 6-5. MeasZ Function Calibration Setup
Set the Calibrator Mainframe in Scope Cal mode at the prompt to connect a 50
Then follow these steps to calibrate MeasZ.
1. Connect the BNC cable to the SCOPE connector. Connect the other end of the BNC
cable to the BNC(f) connector attached to the 50
2. Press the GO ON blue softkey.
3. Enter the actual 50
The Calibrator Mainframe will warn when the entered value is out of
bounds. If this warning occurs recheck the setup and carefully re-enter the
actual resistance insuring proper multiplier (i.e., m, u, n, p). If the warning
still occurs, repair may be necessary.
4. When prompted by the Calibrator Mainframe, disconnect the 50
connect the 1M resistance to the end of the BNC cable.
5. Press the GO ON blue softkey.
6. Enter the actual 1M resistance.
7. When prompted for the first reference capacitor by the Calibrator Mainframe,
disconnect the 1M resistance and leave nothing attached to the end of the BNC
cable.
8. Press the GO ON blue softkey.
9. Enter 0.
10. When prompted for the second reference capacitor by the Calibrator Mainframe,
connect the 50 pF capacitance to the end of the BNC cable.
11. Press the GO ON blue softkey.
12. Enter the actual 50 pF capacitance.
Calibration and Verification of Square Wave Voltage Functions
SC600
Cable
resistance.
Note
5520A-SC600
5520A
CALIBRATOR
NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
HI
1000V
1V PK
RMS
RMS
MAX
MAX
MAX
LO
20V
RMS
MAX
20A
GUARD
TC
20V PK MAX
20V PK MAX
resistance.
resistance and
SC600 Option
6
SCOPE
OUT
20V
150V
PK
MAX
TRIG
20V
SHELLS
PK
NOT
MAX
GROUNDED
yg056f.eps
resistor.
6-27