Ac Voltage, Squarewave Characteristics; Ac Voltage, Trianglewave Characteristics (Typical); Ac Current (Sinewave) Extended Bandwidth Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual
1-29. AC Voltage, Square Wave Characteristics
Risetime
@ 1 kHz
Typical
1 s
1-30. AC Voltage, Triangle Wave Characteristics (typical)
0.3% of p-p value, from 10% to 90% point
1-31. AC Current (Sine Wave) Extended Bandwidth Specifications
Range
All current ranges,
1-28
Settling Time
Overshoot
@ 1 kHz
@ 1 kHz
Typical
Typical
10 s to 1%
2%
of final value
Linearity to 1 kHz
Frequency
330 mA
0.01 Hz to 10 Hz
Duty Cycle
Duty Cycle Uncertainty
Range
1% to 99%,
(0.02% of period + 100 ns), 50% duty cycle
3.3 V p-p,
(0.05% of period + 100 ns), other duty cycles
0.01 Hz to
from 10% to 90%
100 kHz
1% of p-p value, with amplitude 50% of range
1-Year Absolute Uncertainty
tcal
5 C
(% of output
% of range) [2]
5.0
0.5
Aberrations
Max
Current
Resolution
2 digits