3-32. Verifying Phase Accuracy, Volts and AUX Volts

Verify that the 5520A performance is within the limits in Table 3-32, using a precision
phase meter see Figure 3-15.
Table 3-32. Verification Tests for Phase Accuracy, V and V
Output,
Range, Normal
Normal
Output, V
V
3.29999
3.00000
32.9999
30.0000
329.999
50.000
Range, AUX
Frequency
Output
65 Hz
3.29999 V
400 Hz
1 kHz
5 kHz
10 kHz
30 kHz
65 Hz
400 Hz
1 kHz
5 kHz
10 kHz
30 kHz
65 Hz
400 Hz
1 kHz
5 kHz
10 kHz
30 kHz
65 Hz
65 Hz
Calibration and Verification
Performance Verification Tests
Output, AUX
Phase
Lower Limit
3.00000 V
0
60
90
Upper Limit
-0.10
0.10
-0.25
0.25
-0.50
0.50
-2.50
2.50
-5.00
5.00
-10.00
10.00
59.90
60.10
59.75
60.25
59.50
60.50
57.50
62.50
55.00
65.00
50.00
70.00
89.90
90.10
89.75
90.25
89.50
90.50
87.50
92.50
85.00
95.00
80.00
100.00
89.90
90.10
89.90
90.10
3
3-53

Hide quick links: