command:
:SENSe[n]:[CHANnel[n]]:POWer:REFerence:DISPlay
syntax:
:SENSe[n]:[CHANnel[n]]:POWer:REFerence:DISPlay
description:
Takes the input power level value as the reference value.
parameters:
none
response:
none
example:
sens1:pow:ref:disp
affects:
All power meter modules
command:
:SENSe[n]:[CHANnel[n]]:POWer:REFerence:STATe
syntax:
:SENSe[n]:[CHANnel[n]]POWer:REFerence:STATe<wsp><boolean>
description:
Sets the measurement units to relative or absolute units.
parameters:
A boolean value:
response:
none
example:
sens1:pow:ref:stat 1
affects:
All power meter modules
command:
:SENSe[n]:[CHANnel[n]]:POWer:REFerence:STATe?
syntax:
:SENSe[n]:[CHANnel[n]]POWer:REFerence:STATe?
description:
Inquires whether the current measurement units are relative (dB) or absolute
(Watts or dBm).
parameters:
none
response:
A boolean value:
sens1:pow:ref:stat? → 1<END>
example:
affects:
All power meter modules
Introduction to Programming
Measurement Functions – The SENSe Subsystem
0 or OFF: absolute
1 or ON: relative
0: absolute
1: relative