HP 8163A Programming Manual page 87

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe:AUTO
syntax:
SENSe[n]:[CHANnel[n]]:POWer:RANGe:AUTO <wsp><boolean>
description:
Enables or disables automatic power ranging for the module.
If automatic power ranging is enabled, ranging is automatically determined
by the instrument. Otherwise, it must be set by the sens n :pow:rang
command.
parameters:
A boolean value:
response:
none
example:
sens1:pow:rang:auto 1
affects:
All power meter modules
command:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe:AUTO?
syntax:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe:AUTO?
description:
Returns whether automatic power ranging is being used by the module.
parameters:
none
response:
A boolean value:
example:
sens1:pow:rang:auto?
affects:
All power meter modules
Introduction to Programming
Measurement Functions – The SENSe Subsystem
0 or OFF: automatic ranging disabled
1 or ON: automatic ranging enabled
0: automatic ranging is not being used.
1: automatic ranging is being used.
87
1<END>

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8164a

Table of Contents