HP 8163A Programming Manual page 58

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
:STATusn:OPERation:CONDition?
syntax: :STATusn:OPERation:CONDition?
description: Returns the Operational Slot Status Condition Register of slot n.
parameters: none
response: The results for the individual slot events (a 16-bit signed integer value,
where 0 ≤ value ≤ 32767):
Bit
4-15
3
2
1
0
stat1:oper:cond? → +0<END>
example:
command: :STATus:OPERation:ENABle
syntax:
:STATus:OPERation:ENABle<wsp><value>
description:
Sets the bits in the Operational Status Enable Summary Mask (OSESM)
that enable the contents of the OSESR to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR
to affect bit 7 of the Status Byte.
parameters:
The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
response:
none
example:
stat:oper:enab 128
Instrument Setup and Status
Status Reporting – The STATus Subsystem
Mnemonic
Not used
Slot n: Zeroing ongoing
Not used
Slot n: Coherence Control is switched on
Slot n: Laser is switched on
58
Decimal Value
0
8
0
2
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8164a