command:
:STATusn:OPERation[:EVENt]?
syntax: :STATusn:OPERation[:EVENt]?
description: Returns the Operational Slot Status Event Register (OSSER) of slot n.
parameters: none
response: The results for the individual slot events (a 16-bit signed integer value, where 0
≤ value ≤ 32767):
Bit
4-15
3
2
1
0
stat1:oper? → +0<END>
example:
command: :STATus:OPERation:CONDition?
syntax:
:STATus:OPERation:CONDition?
description:
Reads the Operational Status Condition Summary Register.
parameters:
none
response:
The sum of the results for the individual slots (a 16-bit signed integer value,
where 0 ≤ value ≤ 32767):
Bit
5-15
4
3
2
1
0
Every nth bit is the summary of slot n.
stat:oper:cond? → +0<END>
example:
Instrument Setup and Status
Status Reporting – The STATus Subsystem
Mnemonic
Not used
Slot n: Zeroing started
Not used
Slot n: Coherence Control has been switched on
Slot n: Laser has been switched on
Mnemonic
Not used
Summary of slot 4
Summary of slot 3
Summary of slot 2
Summary of slot 1
Summary of slot 0
57
Decimal Value
0
8
0
2
1
Decimal Value
0
16
8
4
2
1