HP 8163A Programming Manual page 67

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command: :SYSTem:PRESet
syntax:
:SYSTem:PRESet
description:
Sets the mainframe and all installed modules to their standard settings. This
command has the same function as the Preset hardkey.
The following are not affected by this command:
• the HP-IB (interface) state,
• the backlight and contrast of the display,
• the interface address,
• the output and error queues,
• the Service Request Enable register (SRE),
• the Status Byte (STB),
• the Standard Event Status Enable Mask (SESEM), and
• the Standard Event Status Register (SESR).
parameters:
none
response:
none
example:
SYST:PRES
command: :SYSTem:TIME
syntax:
:SYSTem:TIME<wsp><hour>,<minute>,<second>
description:
Sets the instrument's internal time.
parameters:
the first value is the hour (0 .. 23),
• the second value is the minute, and
• the third value is the seconds.
response:
none
example:
syst:time 20,15,30
Instrument Setup and Status
Interface/Instrument Behaviour Settings – The SYSTem
Subsystem
67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8164a

Table of Contents