HP 8163A Programming Manual page 54

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
*TST?
syntax:
*TST?
description:
The self-TeST query *TST? makes the instrument perform a self-test and
place the results of the test in the output queue.
If the self-test fails, the results are also put in the error queue. We
recommend that you read self-test results from the error queue.
No further commands are allowed while the test is running. After the self-
test the instrument is returned to the setting that was active at the time the
self-test query was processed.
The self-test does not require operator interaction beyond
sending the *TST? query.
parameters:
none
response:
The sum of the results for the individual tests (a 32-bit signed integer
value, where 0 ≤ value ≤ 4294967296):
Bit
31
5-30
4
3
2
1
0
If 16 is returned, the module in slot 4 has failed.
If 18 is returned, the modules in slots 1 and 4 have failed.
A value of zero indicates no errors.
*TST? → 0<END>
example:
Instrument Setup and Status
IEEE-Common Commands
Mnemonic
Self-test on mainframe
Not used
Self-test on slot 4
Self-test on slot 3
Self-test on slot 2
Self-test on slot 1
Self-test on slot 0
54
Decimal Value
2147483648
0
16
8
4
2
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8164a

Table of Contents