HP 8163A Programming Manual page 114

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:CYCLes?
syntax:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:CYCLes?
[<wsp>MIN|MAX|DEF]
description:
Returns the number of cycles.
parameters:
none
Also allowed are:
response:
The number of cycles as an integer value.
example:
wav:swe:cycl?
affects:
All Tunable Laser modules
command:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:DWELl
syntax:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:DWELl<wsp>
<value>|MIN|MAX|DEF[NS|US|MS|S]
description:
Sets the dwell time.
parameters:
The dwell time as a float value.
If you specify no units in your command, seconds are used as the default.
Also allowed are:
response:
none
example:
wav:swe:dwel 500ms
affects:
All Tunable Laser modules
Introduction to Programming
Signal Generation – The SOURce Subsystem
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but
is half the sum of, the minimum programmable
value and the maximum programmable value
+3<END>
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but
is half the sum of, the minimum programmable
value and the maximum programmable value
114