HP 8163A Programming Manual page 116

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:MODE?
syntax:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:MODE?
description:
Returns the sweep mode.
parameters:
none
response:
STEP:
MAN:
CONT:
example:
wav:swe:mode?
affects:
All Tunable Laser modules
command:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:PMAX?
syntax:
[:SOURce[n]][:CHANnel[n]]:WAVelength:SWEep:PMAX?<wsp>
<start wavelength>,<stop wavelength>
description:
Returns the power to the highest permissible power for the selected wave-
length sweep.
parameters:
start wavelength:
stop wavelength:
response:
The highest permissible power for the selected wavelength sweep as a float
value.
wav:swe:pmax? 1550,157 → +3.5500000E-004<END>
example:
affects:
All Tunable Laser modules
Introduction to Programming
Signal Generation – The SOURce Subsystem
Stepped sweep mode
Manual sweep mode
Continuous sweep mode
STEP<END>
The wavelength at which the sweep starts as a float
value.
The wavelength at which the sweep starts as a float
value.
116

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8164a

Table of Contents