command:
:TRIGger[n][:CHANnel[n]]:OUTPut?
syntax:
:TRIGger[n][:CHANnel[n]]:OUTPut?
description:
Returns the condition that causes an output trigger.
parameters:
none
response:
DISabled:
AVGover:
MEASure:
MODulation:
STFinished:
SWFinished:
SWSTarted:
THReshold:
example:
trig1:outp?
Introduction to Programming
Triggering - The TRIGger Subsystem
Never.
When averaging time finishes.
When a complete measurement (for example a logging
function operation) is finished.
For every leading edge of a digitally-modulated (TTL)
signal
When a sweep step finishes.
When a complete sweep finishes.
When a sweep step starts.
When the threshold condition is met, see
":SENSe[n][:CHANnel[n]]:FUNCtion:THReshold" on
page 84.