command:
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:STABility?
syntax:
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:STABility?
description:
Returns the total time, period time, and averaging time for the stability func-
tion.
parameters:
none
response:
Total time, delay time, and averaging time are float values in seconds.
example:
sens1:func:par:stab?
+1.00000000E-001,+1.00000000E-001<END>
affects:
All power meter modules
command:
:SENSe[n][:CHANnel[n]]:FUNCtion:RESult?
syntax:
:SENSe[n][:CHANnel[n]]:FUNCtion:RESult?
description:
Returns the data array of the last function.
parameters:
none
response:
The last function's data array as a binary block, one measurement value is 4
bytes long in Intel byte order.
example:
sens1:func:res?
affects:
All power meter modules
command:
:SENSe[n][:CHANnel[n]]:FUNCtion:STATe
syntax:
:SENSe[n][:CHANnel[n]]:FUNCtion:STATe<wsp>
LOGGing|STABility|MINMax,STOP|STARt
description:
Enables/Disables the logging, MinMax, or stability function mode.
parameters:
LOGGing:
STABility:
MINMax:
STOP:
STARt:
response:
none
example:
sens1:func:stat
affects:
All power meter modules
Introduction to Programming
Measurement Functions – The SENSe Subsystem
→ returns a data array
logg,star
83
→ +1.00000000E+000,
Logging function
Stability function
MinMax function
Stop function
Start function