HP 8163A Programming Manual page 138

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command: :DISPlay:CONTrast?
syntax:
:DISPlay:CONTrast?
description:
Requests the contrast for the display.
parameters:
none
response:
An integer value in the range 0 to 100
disp:cont? → +50<END>
example:
affects:
HP 8163A Lightwave Multimeter
command: :DISPlay:ENABle
syntax:
:DISPlay:ENABle<wsp><boolean>
description:
Enables or disables the display.
parameters:
A boolean value:
NOTE
response:
none
example:
disp:enab 1
command: :DISPlay:ENABle?
syntax:
:DISPlay:ENABle?
description:
Queries the state of the display.
parameters:
none
response:
A boolean value:
disp:enab? → 1<END>
example:
Instrument Setup and Status
Display Operations – The DISPlay Subsystem
0 – switch off the display
1 – switch on the display
If you press [L
] softkey, the display is enabled automatically.
OCAL
0 – the display is turned off
1 – the display is turned on
138

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8164a

Table of Contents