HP 8163A Programming Manual page 48

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
*ESE
syntax:
*ESE<wsp><value>
0 ≤ value ≤ 255
description:
The standard Event Status Enable command (*ESE) sets bits in the Stand-
ard Event Status Enable Mask (SESEM) that enable the corresponding bits
in the standard event status register (SESR).
The register is cleared:
• at power-on,
• by sending a value of zero.
The register is not changed by the *RST and *CLS commands.
parameters:
The bit value for the register (a 16-bit signed integer value):
Bit
7 (MSB)
6
5
4
3
2
1
0 (LSB)
response:
none
example:
*ESE 21
command:
*ESE?
syntax:
*ESE?
description:
The standard Event Status Enable query *ESE? returns the contents of the
Standard Event Status Enable Mask (see *ESE for information on this
register).
parameters:
none
response:
The bit value for the register (a 16-bit signed integer value).
*ESE? → 21<END>
example:
Instrument Setup and Status
IEEE-Common Commands
Mnemonic
Power On
Not Used
Command Error
Execution Error
Device Dependent Error
Query Error
Not Used
Operation Complete
48
Decimal Value
128
0
32
16
8
4
0
1