Performing quality control (QC) tests
The following procedures must be performed daily, weekly or annually as
indicated.
The reason for performing quality control tests is to determine if any
significant image quality variation or deterioration has occurred which may
require corrective action. Comparing the results of the tests with the reference
values previously established does this.
This procedure allows the operator to take the necessary preventive actions
before any image quality loss can take place.
Performing the daily QC test
This test must be performed every day before any clinical film can be
processed.
Turn on the Drystar 4500 and wait at least for 15 minutes. Refer to
1
the Drystar 4500'
Press the Key-operator key to enter the Key-operator mode.
2
Press the Down key seven times, followed by the Confirm key to select 'QC'.
3
The 'Select input tray' screen appears:
Press the Up/Down arrow keys to select the proper input tray, followed by the
4
Confirm key.
The Drystar 4500 will automatically print the QC Test image.
142
Advanced operation (key-operator mode)
on page 26.
'Switching on
2800E EN 20050215