Film calibration
The film calibration combines Dmax and sensitometry calibration and is
activated as one single procedure. Separate Dmax and sensitometry
calibration steps are not required.
For more general information on maintaining image quality, refer to
'Maintaining image quality and resolving image quality problems'
In the procedure, we will assume a film calibration on the upper input tray. The
procedure for the lower input tray is identical.
Perform steps 1 to 2 of the
1
page
105
to select the 'Calibration' screen.
Press the Confirm key to select 'Film'.
2
The 'Select input tray' screen appears:
Press the Up/Down arrow keys to select the proper input tray, followed by the
3
Confirm key.
The 'Film calibration in progress' screen appears:
The counter indicates the progress of the calibration process. The calibration film is
measured automatically by the built-in MDM (Macro Densitometer). The
Drystar 4500 is being calibrated during the printing process.
In case the film calibration has failed (in this case from upper input tray), the
4
following message will appear.
Press the Confirm key to retry the calibration.
106
Advanced operation (key-operator mode)
'Accessing the Calibration menu'
Failed
Please retry
on page 191.
procedure on
Blinking
Blinking