In case the film calibration has completed successfully, the following message is
6
displayed:
The percentage for the 'Registration compensation' reflects the degree to which the
system had to compensate for incorrect horizontal positioning of the built-in
densitometer (100% compensation = maximal correction).
The percentage for the 'Print head profile compensation' reflects the degree to which
the system had to compensate for density variations in the Print head profile (100%
compensation = maximal correction).
Both values are important for the service technician.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise, the old calibration parameters
will be used.
Check the printed test films.
7
The test films must be similar to the images below (lay-out can differ depending on
the printed format).
Remark: the calibration film must be free of dust particles or any other
artifacts. If this is not the case, you must restart the calibration procedure.
If the printed test film does not look similar to the image above, repeat the
print head profile calibration until the result is satisfactory.
112
Advanced operation (key-operator mode)
Completed succesfully
Blinking
2800E EN 20050215