When the film calibration has completed successfully, three cases are possible:
5
Calibration successful, results OK.
•
The obtained Dmax (x.xx) after calibration is displayed:
Completed succesfully
Calibration successful, but Dmax differs more than 2% from the target
•
value.
The obtained Dmax (x.xx) after calibration is displayed:
Completed succesfully
Calibration successful, but the maximum density has been reached.
•
The obtained Dmax (x.xx) after calibration is displayed:
Completed succesfully
Press the Confirm key to return to the Calibration menu.
6
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise, the old calibration parameters
will be used.
2800E EN 20050215
x.xx
x.xx
x.xx
D
4500 R
RYSTAR
EFERENCE MANUAL
Blinking
Default Densitometer
e.g. Macbeth TR924
Blinking
Default Densitometer
e.g. Macbeth TR924
Blinking
Default Densitometer
e.g. Macbeth TR924
Advanced operation (key-operator mode)
107