Connect the calibration ROM to the Sensor Control-
ler, and then tighten the supplied M2 screw.
Tightening torque: 0.15 N · m max.
Important
• Before connecting or disconnecting the calibration ROM, make sure that the Sensor Controller's power supply is
turned OFF. The Sensor Controller may break down if the calibration ROM is connected while the power is ON.
• Use with the Calibration ROM always connected. If the Calibration ROM is not connected, the error message
"SYSERR" is shown in the main digital and "NO.ROM" is shown in the sub digital, where measurement is not possi-
ble.
"6-1 Error Messages" described in Displacement Sensor ZW-7000 series Confocal Fiber Type Dis-
placement Sensor User's Manual for Communications Settings (Z363-E1-01)
• Be sure to use a Sensor Head and Calibration ROM with the same serial number. A pair with different serial num-
bers cannot operate normally.

Remove the Fiber cable

1
When you remove the fiber connector, make sure to
have the black part of the connector which indicated
the right figure.
Important
If you have the bush section of fiber cable, that part may be out of place.
Make sure to have the fiber connector section.
56
Installation
M2 screw
Calibration ROM
ROM connector
Have the fiber connector (black part)
illustrated under figure, and then pull out
the direction of the arrow.
ZW-7000 User's Manual