Hold clearing conditions
Held values can be cleared by the following operation.
• When FUNC (adjustment) mode/RUN (operation) mode is switched
• When hold RESET is entered (32-pole extension connector,
• Hold RESET command is entered.
(Example) Peak
The mode is switched between
the FUNC mode and RUN mode
FUNC mode
Clamp level
(Example) Auto Peak
The mode is switched between
the FUNC mode and RUN mode
FUNC mode
Clamp level
Important
If an abnormal measured value is obtained during sampling, the hold value is cleared. To not clear the hold value even
if an abnormal measured value is obtained, set "KEEP" as the non-measurement setting.
"2-1 Parallel I/O connection" described in Displacement Sensor ZW-7000 series Confocal Fiber Type
Displacement Sensor User's Manual for Communications Settings (Z363-E1-01)
ZW-7000 User's Manual
RUN mode
When RESET is executed
(32-pole extension connector/key input/command).
Sampling
Sampling
RUN mode
When RESET is executed
(32-pole extension connector/key input/command).
key)
When the hold value is reset,
the clamp level is output until
the next hold value is applied.
Hold value
Measurement
value
Hold value
Measurement
value
Setting the Output Conditions
123