3
After internal logging ends, click the [Save to
file (Sensor → PC)] to output the data to a
file.
A CSV format file in the following format is
output.
LoggingMode
SamplePeriod
SamplingCycleTime
Index
0
1
2
3
4
5
Set the conditions for the internal logging using LOGGING input signal of the parallel I/O.
 Explorer pane
→ Edit pane
→ Sensor settings window
1
Set the logging conditions for internal logging.
There are three settings: Number of logging, number of logging decimation, and overwrite mode.
Item
Internal logging
setting
Exceeding the maximum number of logging results in the following behavior according to the setting of
overwrite mode:
• With overwrite mode turned OFF
The internal logging stops.
• With overwrite mode turned ON
The internal logging continues.
The oldest logging data is overwritten.
Note
When continue to logging after starting logging until stop logging, set the as below:
• Logging number: 0
• Overwriting mode: ON
156
Performing internal logging
Internal
1
0.2
OUT1
1.21314
1.22098
0.12334
-0.1211
-1.23456
-1.22222
: [System] (double-click)
: [Sensor Settings] icon (
: [Internal logging setting]
Setting item
Setting value
Number of
0 to 2000000
logging data
(default value: 100000)
Internal logging
0 to 1000
settings
(default value: 1)
Overwrite mode
OFF (default value)/ON
OUT2
OUT3
1.21314
1.21314
1.22098
1.22098
0.12334
0.12334
-0.1211
-0.1211
-1.23456
-1.23456
-1.22222
-1.22222
)
Description
Set the number of data to save for each internal logging
output data.
Set the interval of saving internal logging data.
Set whether to make the internal logging mode run in
overwrite mode.
OUT4
1.21314
1.22098
0.12334
-0.1211
-1.23456
-1.22222
ZW-7000 User's Manual