Extending fiber cable
To extend the fiber cable on the Sensor Head, use an extension fiber cable and connecting adapter.
Extension fiber cable (option)
Model
ZW-XF7002R
ZW-XF7005R
Connecting adapter (option)
Model
ZW-XFCM
Important
• External fiber cable has the direction. Make sure to connect the name plate side to the Sensor Controller.
• When connect in the opposite direction, can not be transfer the optical signal and the correctly measurement.
• If you connect having fiber connector, cannot be connected correctly.
• Additionally, if force to connect the standard fiber cable and extension fiber cable in the opposite direction, it may be
caused failure.
• For how to use the supplied strap, refer to the following; Reference mark p.26.
60
Installation
Sensor Head
fiber cable
When connecting the fiber cable, grasp by the bushing part
of fiber. Align the convex section of fiber cable to the groove
of the adapter and insert it until there is a clicking sound.
Fiber connector
Convex section
Fiber bush part
Length
2 m
5 m
Connection adaptor
ZW-XFCM
Extension fiber cable
ZW-X70□□R:
(2 m/5 m)
Groove
Nameplate