5
Commands by Subsystem

:TRIGger:DURation Commands

Table 58 :TRIGger:DURation Commands Summary
Command
:TRIGger:DURation:GRE
aterthan <greater
than time>[suffix]
(see
page
377)
:TRIGger:DURation:LES
Sthan <less than
time>[suffix] (see
page
378)
:TRIGger:DURation:PAT
Tern <value>, <mask>
(see
page
379)
:TRIGger:DURation:QUA
Lifier <qualifier>
(see
page
380)
:TRIGger:DURation:RAN
Ge <greater than
time>[suffix], <less
than time>[suffix]
(see
page
381)
376
Query
:TRIGger:DURation:GRE
aterthan? (see
page
377)
:TRIGger:DURation:LES
Sthan? (see
page
378)
:TRIGger:DURation:PAT
Tern? (see
page
379)
:TRIGger:DURation:QUA
Lifier? (see
page
380)
:TRIGger:DURation:RAN
Ge? (see
page
381)
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
Options and Query Returns
<greater than time> ::=
floating-point number from 5 ns
to 10 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
<less than time> ::=
floating-point number from 5 ns
to 10 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
<value> ::= integer or <string>
<mask> ::= integer or <string>
<string> ::= ""0xnnnnnn"" n ::=
{0,..,9 | A,..,F}
<qualifier> ::= {GREaterthan |
LESSthan | INRange | OUTRange |
TIMeout}
<greater than time> ::= min
duration from 10 ns to
9.99 seconds in NR3 format
<less than time> ::= max duration
from 15 ns to 10 seconds in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}