:TRIGger:UART:SOURce:TX

Command Syntax
:TRIGger:UART:SOURce:TX <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for the MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:UART:SOURce:TX command controls which signal is used as
the Tx source by the serial decoder and/or trigger when in UART mode.
Query Syntax
:TRIGger:UART:SOURce:TX?
The :TRIGger:UART:SOURce:TX? query returns the current source for the
UART Tx signal.
Return Format
<source><NL>
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 360
":TRIGger:UART:TYPE"
":TRIGger:UART:BITorder"
on page 354
on page 440