:TRIGger:UART:TYPE

Command Syntax
:TRIGger:UART:TYPE <value>
<value> ::= {RSTArt | RSTOp | RDATa | RD1 | RD0 | RDX | PARityerror
The :TRIGger:UART:TYPE command selects the UART trigger type.
When one of the RD or TD types is selected, the :TRIGger:UART:DATA and
:TRIGger:UART:QUALifier commands are used to specify the data value
and comparison operator.
The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and
alert bit values) are only valid when a 9- bit width has been selected.
Query Syntax
:TRIGger:UART:TYPE?
The :TRIGger:UART:TYPE? query returns the current UART trigger data
value.
Return Format
<value><NL>
<value> ::= {RSTA | RSTO | RDAT | RD1 | RD0 | RDX | PAR | TSTA |
See Also
440
(see
page
586)
| TSTArt | TSTOp | TDATa | TD1 | TD0 | TDX}
TSTO | TDAT | TD1 | TD0 | TDX}
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 360
":TRIGger:UART:DATA"
":TRIGger:UART:QUALifier"
":TRIGger:UART:WIDTh"