:OPERegister:CONDition (Operation Status Condition Register)

Query Syntax
:OPERegister:CONDition?
The :OPERegister:CONDition? query returns the integer value contained in
the Operation Status Condition Register.
Table 38 Operation Status Condition Register
Bit
15-13
12
11
10-6
5
4
3
2-0
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
Name
Description
---
---
HWE
Hardware Event
OVLR
Overload
---
---
Wait
Wait Trig
Trig
---
---
Run
Running
---
---
Commands by Subsystem
When Set (1 = High = True), Indicates:
(Not used.)
A hardware event has occurred..
A 50Ω input overload has occurred.
(Not used.)
The trigger is armed (set by the Trigger Armed
Event Register (TER)).
(Not used.)
The oscilloscope is running (not stopped).