Measure:pwidth - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:MEASure:PWIDth

Command Syntax
:MEASure:PWIDth [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:PWIDth command installs a screen measurement and starts
the positive pulse width measurement. If the optional source parameter is
specified, the current source is modified.
This command is not available if the source is FFT (Fast Fourier Transform).
N O T E
Query Syntax
:MEASure:PWIDth? [<source>]
The :MEASure:PWIDth? query measures and outputs the width of the
displayed positive pulse closest to the trigger reference. Pulse width is
measured at the midpoint of the upper and lower thresholds.
IF the edge on the screen closest to the trigger is falling:
THEN width = (time at trailing falling edge - time at leading rising edge)
ELSE width = (time at leading falling edge - time at leading rising edge)
Return Format
<value><NL>
<value> ::= width of positive pulse in seconds in NR3 format
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
586)
"Introduction to :MEASure Commands"
":MEASure:SOURce"
":MEASure:NWIDth"
":MEASure:PERiod"
on page 272
on page 257
on page 279
on page 269
Commands by Subsystem