Operation Troubleshooting - Siemens SITRANS LR560 Operating Instructions Manual

Radar transmitters ma/hart
Hide thumbs Also See for SITRANS LR560:
Table of Contents

Advertisement

Diagnosing and troubleshooting

10.3 Operation troubleshooting

10.3
Operation troubleshooting
Operating symptoms, probable causes, and resolutions.
Symptom
Cause
Display shows
level or target is out of range
Display shows
material build-up on antenna
Display shows
location or aiming:
Display shows
antenna malfunction:
Reading does not
SITRANS LR560 processing
change, but the level
wrong echo, for example, vessel
does
wall, or structural member
Measurement is con-
sistently off by a con-
stant amount
Screen blank
power error
Reading erratic
echo confidence weak
material filling
Reading response
FILL RATE/MIN (2.4.2.)
slow
(Page 115) setting incorrect
160
poor installation
flange not level
Auto False Echo Suppression
may be incorrectly applied
temperature too high
physical damage
setting for LOW
CALIBRATION POINT
(2.3.1.) (Page 113) not cor-
rect
setting for SENSOR OFFSET
(2.3.3.) (Page 114) not cor-
rect
Action
check specifications
check LOW CALIBRATION POINT (2.3.1.) (Page 113).
increase CONFIDENCE (2.7.6.1) (Page 126).
Use the air purge feature to clean the antenna
re-locate SITRANS LR560.
check to ensure nozzle is vertical
ensure end of antenna protrudes from end of nozzle
review Echo processing (Page 180)
ensure Auto False Echo Suppression Range is set correct-
ly
check temperature in MAXIMUM MEASURED VALUE
(3.4.2.) (Page 131).
relocate
re-locate SITRANS LR560
check nozzle for internal burrs or welds
rotate device 90°
use AUTO FALSE ECHO SUPPRESSION (2.8.1)
(Page 126) if necessary.
check distance from sensor reference point to LOW
CALIBRATION POINT (2.3.1.) (Page 113)
check SENSOR OFFSET (2.3.3.) (Page 114)
check nameplate rating against voltage supply
check power wiring or source
refer to CONFIDENCE (2.7.6.1) (Page 126).
use AUTO FALSE ECHO SUPPRESSION (2.8.1)
(Page 126) and AUTO FALSE ECHO SUPPRESSION
RANGE (2.8.2) (Page 127)
use foam deflector or stillpipe
Re-locate SITRANS LR560.
increase measurement response if possible
Operating Instructions, 06/2017, A5E34647946-AD