S: 94
Device error. Possible causes:
1. Voltage low at the LR560
2. Hardware defect
S: 95
Device error.
to
NOTE: Fault text and icons appear only on the local
S:112
display.
SITRANS LR560 (mA/HART)
Operating Instructions, 06/2017, A5E34647946-AD
Diagnosing and troubleshooting
Corrective Action
For 1: Confirm voltage is within specifications and
modify wiring or increase source voltage according-
ly, see Typical connection drawing (Page 190).
For 2: Repair required: contact your local Siemens
representative.
Repair required: contact your local Siemens repre-
sentative.