Diagnosing and troubleshooting
10.1
Device status icons
Icon
Priority
Meaning
Level
1
Maintenance alarm
•
Measurement values are not valid
•
2
Maintenance warning: maintenance demanded immediately
•
Measured signal still valid
•
3
Maintenance required
•
Measured signal still valid
•
1
Process value has reached an alarm limit
•
2
Process value has reached a warning limit
•
3
Process value has reached a tolerance limit
•
1
Configuration error
•
Device will not work because one or more parameters/components is incorrectly configured
•
2
Configuration warning
•
Device can work but one or more parameters/components is incorrectly configured
•
3
Configuration changed
•
Device parameterization not consistent with parameterization in project. Look for info text.
•
1
Manual operation (local override)
•
Communication is good; device is in manual mode.
•
2
Simulation or substitute value
•
Communication is good; device is in simulation mode or works with substitute values.
•
3
Out of operation
•
Communication is good; device is out of action.
•
No data exchange
•