5502A
Operators Manual
AC Current (Sine Wave) (cont.)
Range
29 to 329.99 μA
330 μA to
3.29999 mA
3.3 to
32.9999 mA
33 to 329.999 mA
330 mA to
2.99999 A
3.3 A to 20.5 A
[1]
Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1-3. The current may be
provided 60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is
the output current in amps. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the
5502A is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
Under those conditions, the allowable "on" time indicated by the formula and Figure 1-3 is achieved only after the 5502A is
outputting currents <5 A for the "off" period first.
For currents >11 A, Floor specification is 4000 μA within 30 seconds of selecting operate. For operating times >30 seconds,
[2]
the floor specification is 2000 μA.
For currents >11 A, Floor specification is 1000 μA within 30 seconds of selecting operate. For operating times >30 seconds,
[3]
the floor specification is 5000 μA.
[4]
Subject to compliance voltages limits.
29 to 329.99 μA
0.33 to 3.29999 mA
3.3 to 32.9999 mA
33 to 329.999 mA
0.33 to 2.99999 A
3 to 20.5 A
[1]
Subject to specification adder for compliance voltages greater than 1 V rms.
1-16
Absolute Uncertainty, tcal
±5 °C ±(% of output + μA)
Frequency
90 Day
10 to 100 Hz
0.20 + 0.2
100 Hz to 1 kHz
0.50 + 0.5
10 to 100 Hz
0.20 + 0.3
100 Hz to 1 kHz
0.50 + 0.8
10 to 100 Hz
0.07 + 4
100 Hz to 1 kHz
0.18 + 10
10 to 100 Hz
0.07 + 40
100 Hz to 1 kHz
0.18 + 100
10 to 100 Hz
0.10 + 200
100 to 440 Hz
0.25 + 1000
45 to 100 Hz
0.10 + 2000
[1]
100 to 440 Hz
0.80 + 5000
Range
Max Distortion and
Noise 10 Hz to
100 kHz BW ±(% of
1 Year
output + floor)
LCOMP On
0.25 + 0.2
0.05 + 1.0 μA
0.60 + 0.5
0.25 + 0.3
0.15 + 1.5 μA
0.60 + 0.8
0.08 + 4
0.20 + 10
0.08 + 40
0.20 + 100
0.12 + 200
0.30 + 1000
0.25 + 500 μA
[2]
[2]
0.12 + 2000
[3]
[3]
1.00 + 5000
Resolution μA
0.01
0.01
0.1
1
10
100
Max Inductive Load
0.1 + 1.0 μA
0.06 + 1.5 μA
0.15 + 5 μA
400 μH
0.05 + 5 μA
0.15 + 50 μA
0.05 + 50 μA
0.2 + 500 μA
0.1 + 0 μA
400 μH
0.5 + 0 μA
Max Compliance Voltage V rms
7
7
5
5
4
3
[4]
[1]