Example:
OPER? shows 1
Shows 1 when the Calibrator is in operate.
x
IEEE-488
*OPT?
(Options command) Shows a list of the installed hardware and software options.
Responses:
<option string>,<option string>,... (options list, separated by commas)
0 (no options are installed)
Example:
*OPT? shows SC600
Shows SC600 when the Oscilloscope Calibration Option is installed.
x
x
IEEE-488
OUT
(Output command) Sets the output of the Calibrator and establishes a new reference
point for the error mode. If only one amplitude is supplied, the Calibrator sources a single
output. If two amplitudes are supplied, the Calibrator sources two outputs. The second
amplitude will be sourced at the AUX terminals for dual voltage outputs. If the frequency
is not supplied, the Calibrator will use the frequency that is currently in use.
To source or measure a temperature, select the desired sensor and sensor parameters first.
(See the TSENS_TYPE, RTD_*, and TC_* commands.)
To source a signal with the Calibrator scope options, refer to the SCOPE command.
If you change the frequency of an ac function and the harmonic output is not explicitly
set at the same time with the HARMONIC command, the harmonic will be set to 1.
Use multipliers e.g., k, M, μ with the OUT command, as desired.
Parameters: <value> V
<value> DBM
<value> V, <value> Hz
<value> DBM, <value> Hz
<value> A
<value> A, <value> Hz
<value> OHM
<value> F
<value> CEL
<value> FAR
<value> HZ
<value> V, <value> A
<value> V, <value> A, <value> HZ
<value> V, <value> V
<value> V, <value> V, <value> HZ
<value>
Examples:
OUT 15.2 V
OUT 20 DBM
OUT 10 V, 60 Hz
OUT 10 DBM, 50 HZ
OUT 1.2 MA
OUT 1 A, 400 HZ
x
x
RS-232
Sequential
x
RS-232
Sequential
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
Volts dc or update volts ac
Volts ac dBm update
Volts ac or volts dc with 0 Hz
Volts ac in dBm
Current dc or update current ac
Current ac
Resistance
Capacitance
Temperature (Celsius)
Temperature (Fahrenheit)
Update frequency
Power dc or update power ac
Power ac
Dual volts dc or update dual ac
Dual volts ac in volts
For single output, changes
amplitude keeping unit and
frequency the same.
(volts; 15.2 V @ same frequency)
(volts; 20 dBm @ same frequency)
(volts ac; 10 V @ 60 Hz)
(volts ac; 10 dBm @ 50 Hz)
(current; 1.2 mA @ same frequency)
(current ac; 1 A @ 400 Hz)
6
Remote Commands
Commands
6-25