The Fan Filtercovers the air intake to keep dust and contamination out of the chassis air baffles.
The fan supplies a constant cool air flow to the inner chassis. Instructions for fan filter maintenance
are in Chapter 7.
The CALIBRATION NORMAL/ENABLE slide switch is used to write enable and disable the
nonvolatile memory that stores calibration constants. Change to ENABLE to let changes be written
into memory, and change to NORMAL to protect data in memory from being overwritten. The
switch is recessed to let it to be covered with a calibration sticker to guarantee calibration integrity.
The SERIAL 2 TO UUT connector is used to transmit and receive RS-232 serial data between the
Calibrator and a UUT. Chapter 6, "Remote Commands" tells you how to use the RS-232 serial
interface for UUT communications.
The SERIAL 1 FROM HOST connector is for remote control of the Calibrator and for transmitting
internal-constant RS-232 serial data to a printer, monitor, or host computer. Chapter 5, "Remote
Operation" tells you how to use the RS-232 serial interface for remote control.
The IEEE-488 connectorError! Bookmark not defined. is a standard parallel interface for
operating the Calibrator in remote control as a Talker/Listener on the IEEE-488 bus. Refer to
Chapter 5, "Remote Operation" for bus connection and remote programming instructions.
3-10
CAUTION
FOR FIRE PROTECTION REPLACE ONLY
MAINS SUPPLY
FUSE
WITH A 250V FUSE OF INDICATED RATING
100V/120V
T5.0A 250V
(SB)
47Hz
/63Hz
220V/240V
T2.5A 250V
(SB)
600VA MAX
CHASSIS
GROUND
7
Figure 3-2. Rear-Panel Features
Table 3-2. Rear-Panel Features
1
WARNING:
TO AVOID PHYSICAL INJURY, INSURE THAT THE FILTER
IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRUMENT
WARNING:
TO AVOID ELECTRIC SHOCK GROUNDING
CONNECTOR IN POWER CORD MUST BE CONNECTED
6
2
3
4
NORMAL
ENABLE
INSTALLED
OPTIONS
CALIBRATION
- SC600
- SC1100
- PQ
SERIAL 2
TO UUT
FLUKE CORPORATION
EVERETT WA, USA
SERIAL 1
FROM HOST
NO INTERNAL USER SERVICEABLE
PARTS. REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL
IEEE-488
LR65268C
(LEM CERTIFIED)
10MHZ
IN
OUT
TO CLEAN THE FILTER:
-UNPLUG INSTRUMENT
5V
PK - PK
-REMOVE FILTER
-FLUSH WITH SOAPY WATER
MAX
-DRY BEFORE REINSTALLATION
5
gjh011.eps