Fluke 5502A Operator's Manual page 249

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

Table 7-6. Verification Tests for Resistance (cont.)
Range
109.999 kΩ
109.999 kΩ
329.999 kΩ
329.999 kΩ
329.999 kΩ
1.09999 MΩ
1.09999 MΩ
3.29999 MΩ
3.29999 MΩ
3.29999 MΩ
10.9999 MΩ
10.9999 MΩ
32.9999 MΩ
32.9999 MΩ
32.9999 MΩ
109.999 MΩ
109.999 MΩ
329.999 MΩ
329.999 MΩ
1100.00 MΩ
1100.00 MΩ
1100.00 MΩ
Table 7-7. Verification Tests for AC Voltage (Normal)
Range
32.999 mV
3.000 mV
32.999 mV
3.000 mV
32.999 mV
30.000 mV
32.999 mV
30.000 mV
32.999 mV
30.000 mV
32.999 mV
30.000 mV
32.999 mV
30.000 mV
32.999 mV
30.000 mV
Output
33.000 kΩ
109.000 kΩ
119.000 kΩ
190.000 kΩ
300.000 kΩ
0.33000 MΩ
1.09000 MΩ
1.19000 MΩ
1.90000 MΩ
3.00000 MΩ
3.3000 MΩ
10.9000 MΩ
11.9000 MΩ
19.0000 MΩ
30.0000 MΩ
33.000 MΩ
109.000 MΩ
119.000 MΩ
290.000 MΩ
400.00 MΩ
640.00 MΩ
1090.00 MΩ
Output
Frequency
45 Hz
10 kHz
9.5 Hz
10 Hz
45 Hz
1 kHz
10 kHz
20 kHz
Lower Limit
32.9971 kΩ
108.9910 kΩ
118.9872 kΩ
189.9809 kΩ
299.9710 kΩ
0.329961 MΩ
1.089878 MΩ
1.189839 MΩ
1.899761 MΩ
2.999640 MΩ
3.29846 MΩ
10.89504 MΩ
11.88857 MΩ
18.98325 MΩ
29.99750 MΩ
32.8650 MΩ
108.5610 MΩ
118.4240 MΩ
288.7400 MΩ
394.700 MΩ
631.820 MΩ
1076.420 MΩ
Lower Limit
2.977 mV
2.977 mV
28.350 mV
29.944 mV
29.956 mV
29.956 mV
29.956 mV
29.944 mV
Maintenance
Performance Tests
Upper Limit
33.0028 kΩ
109.0089 kΩ
119.0127 kΩ
190.0191 kΩ
300.0290 kΩ
0.330038 MΩ
1.090121 MΩ
1.190160 MΩ
1.900239 MΩ
3.000360 MΩ
3.30153 MΩ
10.90495 MΩ
11.91142 MΩ
19.01675 MΩ
30.02500 MΩ
33.1350 MΩ
109.4390 MΩ
119.5760 MΩ
291.2600 MΩ
405.300 MΩ
648.180 MΩ
1103.580 MΩ
Upper Limit
3.022 mV
3.022 mV
31.650 mV,
30.056 mV
30.044 mV
30.044 mV
30.044 mV
30.056 mV
7
7-11