Sc600 Oscilloscope Calibration Option Specifications; Voltage Function Specifications; Edge Specifications - Fluke 5502A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

5502A
Operators Manual

SC600 Oscilloscope Calibration Option Specifications

These specifications apply only to the SC600 Oscilloscope Calibration Option. General
specifications that apply to the 5502A Calibrator can be found in Chapter 1. The
specifications are correct given the 5502A is operated with the conditions specified in
Chapter 1, and has completed a warm-up period of no less than two times the length of
time the Calibrator was powered off, to a maximum of 30 minutes.

Voltage Function Specifications

Voltage Function
Amplitude Qualities
Range
Resolution
Adjustment Range
1-Year Absolute Uncertainty,
tcal ±5 °C
Sequence
Square Wave Frequency Qualities
Range
1-Year Absolute Uncertainty,
tcal ±5 °C
Typical aberration within 4 μs from
50 % of leading/trailing edge
[1]
Selectable positive or negative, zero referenced square wave.
For square wave frequencies above 1 kHz, ± (0.25 % of output + 40 μV).
[2]

Edge Specifications

Rise Time
Amplitude Range (p-p)
Resolution
Adjustment Range
Sequence Values
Frequency Range
Typical Jitter, edge to trigger
Leading Edge Aberrations
Typical Duty Cycle
Tunnel Diode Pulse Drive
[1]
Above 2 MHz rise time specification <350 ps
[2]
All edge aberration measurements made with Tektronix 11801 mainframe with SD26 input module.
9-4
50 Ω Load
0 to ±6.599 V
Range
1 to 24.999 mV
25 to 109.99 mV
110 mV to 2.1999 V
2.2 to 10.999 V
11 to 130 V
±(0.25 % of output
+ 40 μV)
Edge Qualities into 50 Ω Load
≤300 ps
4.5 mV to 2.75 V
4 digits
±10 % around each sequence value
(indicated below)
5 mV, 10 mV, 25 mV, 50 mV, 60 mV, 80
mV, 100 mV, 200 mV, 250 mV, 300 mV,
500 mV, 600 mV, 1 V, 2.5 V
[1]
900 Hz to 11 MHz
<5 ps (p-p)
within 2 ns from 50 % of rising edge
2 to 5 ns
[2]
5 to 15 ns
after 15 ns
45 % to 55 %
Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 to 100 V p-p.
DC Signal
1 MΩ Load
0 to ±130 V
±6.599 V p-p
Resolution
1 μV
10 μV
100 μV
1 mV
10 mV
Continuously adjustable
± 0.05 % of output
±(0.25 % of output
+ 40 μV)
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
10 Hz to 10 kHz
±(2.5 ppm of setting)
<(0.5 % of output + 100 μV)
[1]
Square Wave Signal
50 Ω Load
1 MΩ Load
±1 mV to
±1 mV to
±130 V p-p
±(0.1 % of
+ 40 μV)
output + 40 μV)
1-Year Absolute Uncertainty,
tcal ± 5 °C
(+0 ps / -100 ps)
±(2 % of output + 200 μV)
±(2.5 ppm of setting)
<(3 % of output + 2 mV)
<(2 % of output + 2 mV)
<(1 % of output + 2 mV)
<(0.5 % of output + 2 mV)
[2]