Key in a Value

The subsequent example is for use in the LEVSINE mode. To key a specific value
directly into the Calibrator from its front panel:
1. Key in the value you want to record, including the units and prefixes. For example, to
record 120 mV push     . The control display shows:
Units and prefixes printed in red in the upper-left corner of the keys are
accessed through the  key. For example, to record 200
     .
If you make an error, push  to clear the Control Display and return to the menu.
2. Push  to activate the value and move it to the Output Display.
Other settings in the display will remain unaltered unless you key in an entry and
specify the units for that setting.

Adjust Values with the Rotary Knob

To adjust values in the Output Display with the rotary knob:
1. Turn the rotary knob. A cursor shows in the output display under the lowest digit and
begins changing that digit. If you wish to place the cursor in the field without
changing the digit, push , see below.
2. To move the cursor between the voltage and frequency fields, push , see below.
3. Use  and  to move the cursor to the digit you want to change.
SC600 Oscilloscope Calibration Option
Note
Start the SC600 Option
gl002i.eps
μ
s, push 
gjh038.eps
gjh039.eps
9
9-9