You can record the equivalent frequency as an alternative to the time
marker value. For example, you can record 5 MHz for 200 ns.
3. Set the time base of your oscilloscope to show 10 time markers. The time markers
must align with the oscilloscope divisions, as shown in the example below.
For an accurate indication, align the peaks of the signal with the horizontal center
axis.
4. Do this procedure again for all time marker values recommended for your
oscilloscope. Do this procedure again for digital and analog mode as necessary. It can
be necessary, for some oscilloscopes, to change the magnification while you calibrate
in analog mode.
5. Push  to remove the input signal.

Test the Trigger SC600 Option

The oscilloscope can trigger on different waveforms. This can be tested with the wave
generator. When the wave generator is used, a square, sine, or triangle wave is
transmitted and the wave output impedance, offset, and voltage can be changed to test the
trigger function at different levels.
The wave generator must not be used to verify the accuracy of your
oscilloscope.
SC600 Oscilloscope Calibration Option
Note
Peaks are aligned
with center axis
Note
Test the Trigger SC600 Option
gl011i.eps
9
9-23