5. Push a multiplier key (preceded with the  key) for the necessary output. For
6. Push .
7. The Control Display shows the amplitude of your capacitance entry. For example,
8. Push . The Calibrator erases your entry from the Control Display and copies
9. Push  to activate the Calibrator output.
The softkey in the Control Display named COMP lets you select one of three
lead-compensation settings (see below).
4-36
example, push  then  for μF. The other multiplier keys include  for pF
and  for nF.
123.45 μF (see below).
it into the Output Display (below is typical).
COMP (Compensation). Applies 2-wire compensation or turns compensation off.
Compensation refers to methods of connection the Calibrator to the UUT to cancel
out test lead resistance (NOT capacitance). Compensation is available for
capacitances of 110 nF and above. This softkey does not function in less than 110 nF.
See "Four-Wire versus Two-Wire Connections" in this chapter.
nn099f.eps
nn100f.eps
nn101f.eps