Select Line Frequency - Fluke 5502A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

Select Line Frequency

The calibrator is shipped from the factory for nominal operation at 60 Hz line frequency.
If you use 50 Hz line voltage, re-configure the Calibrator for optimal performance at
50 Hz. To do this:
1. From the front panel, go into SETUP, INSTMT SETUP, OTHER SETUP.
2. Push the softkey below MAINS to change the selection to 50 Hz.
3. Store the change.
After a correct instrument warmup (on for 30 minutes or longer), you must zero the
complete instrument again. See the section on "Zero the Calibrator" in Chapter 4.
Changing Line Fuse
Changing Line
Voltage
Figure 2-1. Access the Fuse and Select Line Voltage
Prepare for Operation

Select Line Frequency

Line Voltage
Indicator
0 V
( S
B )
1 2
0
2
1 2
0
gjh004.eps
2-5