Fluke 5502A Operator's Manual page 252

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

5502A
Operators Manual
Range
329.999 mV
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
5.00000 V
5.00000 V
5.00000 V
5.00000 V
5.00000 V
5.00000 V
[1]
Set the NORMAL output to 300 mV.
Range
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
329.99 μA
3.2999 mA
3.2999 mA
3.2999 mA
7-14
Table 7-8. Verification Tests for AC Voltage (AUX) (cont.)
[1]
Output, AUX
Frequency
300.000 mV
30 kHz
3.00000 V
9.5 Hz
3.00000 V
10 Hz
3.00000 V
45 Hz
3.00000 V
1 kHz
3.00000 V
5 kHz
3.00000 V
10 kHz
3.00000 V
30 kHz
5.00000 V
9.5 Hz
5.00000 V
10 Hz
5.00000 V
45 Hz
5.00000 V
1 kHz
5.00000 V
5 kHz
5.00000 V
10 kHz
Table 7-9. Verification Tests for AC Current
Output
Frequency
33.00 μA
1 kHz
33.00 μA
10 kHz
33.00 μA
30 kHz
190.00 μA
45 Hz
190.00 μA
1 kHz
190.00 μA
10 kHz
190.00 μA
30 kHz
329.00 μA
10 Hz
329.00 μA
45 Hz
329.00 μA
1 kHz
329.00 μA
5 kHz
329.00 μA
10 kHz
329.00 μA
30 kHz
0.3300 mA
1 kHz
0.3300 mA
5 kHz
0.3300 mA
30 kHz
Lower Limit
Upper Limit
287.100 mV
312.900 mV
2.835 V
3.165V
2.99505 V
3.00495 V
2.99745 V
3.00255 V
2.99745 V
3.00255 V
2.99410 V
3.00590 V
2.98960 V
3.01040 V
2,87720 V
3.12280 V
4.72500 V
5.27500 V
4.99205 V
5.00795 V
4.99605 V
5.00395 V
4.99605 V
5.00395 V
4.99110 V
5.00890 V
4.98360 V
5.01640 V
Lower Limit
Upper Limit
32.87 μA
33.13 μA
32.60 μA
33.40 μA
32.20 μA
33.80 μA
189.71 μA
190.29 μA
189.71 μA
190.29 μA
188.66 μA
191.34 μA
187.32 μA
192.68 μA
328.37 μA
329.63 μA
328.57 μA
329.43 μA
328.57 μA
329.43 μA
328.03 μA
329.97 μA
326.83 μA
331.17 μA
324.65 μA
333.35 μA
0.3296 mA
0.3304 mA
0.3293 mA
0.3307 mA
0.3268 mA
0.3332 mA

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents