Multipoint calibration

3.1.9.1

Enable multipoint calibration

Description
Enable or disable the multipoint calibration.
Setting
Default
Related
Path
3.1.9.2

Asymmetric calibration table

Description
Note
Access level: Expert user only
Only available for LUILOCAL_CURRENT_LUI_ACCESS_LEVEL == SUP AND
PID_ProtectCalibrationTable == 0 AND PID_EnableUserCalibrationTable == 1
Usage of the calibration table:
● Unidirectional: All 20 calibration points are used for positive flow calibration and are then
mirrored for negative flow.
● Bidirectional: The first 10 calibration points are used for positive flow calibration and the
10 last calibration points are used for negative flow calibration.
Setting
Default
Related
Path
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
Setup → Sensor settings → User calibration → Path 1 offset
Disabled; Enabled
Disabled
Setup → Sensor settings → Multipoint calibration
Unidirectional; Bidirectional
Unidirectional
Setup → Sensor settings → Asymmetric calibration table
Setup
3.1 Sensor settings