Parameter Assignment; Multipoint Calibration - Siemens SITRANS F Operating Instructions Manual

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Parameter assignment

8.1

Multipoint calibration

The transmitter provides an additional 20 point piecewise linear calibration table with user
selectable input value: volume flow, reynolds number or fluid sound velocity. The table can be
configured for either unidirectional or bidirectional calibration.
Note
Ascending order by reynolds number
The datapoints must be entered in ascending order by reynolds number.
SITRANS FST020 IP65 (NEMA 4X)
Operating Instructions, 05/2022, A5E41425845-AC
8
57

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fst020 ip65 nema 4x

Table of Contents